EMC information
117
CLV-190 INSTRUCTION MANUAL
App.
Guidance and manufacturer’s declaration — Electromagnetic
emissions
This model is intended for use by medical personnel in hospital environments and for use in the
electromagnetic environment specified below. The customer or the user of this model should
assure that it is used in such an environment.
EMC information
Emission standard Compliance Guidance
RF emissions
CISPR 11
Group 1 This instrument uses RF (Radio Frequency) energy only for its
internal function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
Radiated emissions
CISPR 11
Class B This instrument’s RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Main terminal
conducted emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Class A This instrument’s harmonic emissions are low and are not likely to
cause any problem in the typical commercial power supply connected
to this instrument.
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies This instrument stabilizes its own radio variability and has no effect
such as flicker in lighting apparatus.

Table of Contents

Other manuals for Olympus EVIS EXERA III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus EVIS EXERA III and is the answer not in the manual?

Olympus EVIS EXERA III Specifications

General IconGeneral
ModelEVIS EXERA III
CategoryMedical Equipment
TypeEndoscopy System
ManufacturerOlympus
System TypeVideo Endoscopy System
Video ProcessorCV-190
HD OutputYes
Power Requirements100-240 V AC, 50/60 Hz
Image QualityHDTV
Compatible EndoscopesEVIS EXERA III series endoscopes
Image EnhancementNBI (Narrow Band Imaging)
Light SourceXenon Lamp

Related product manuals