3.3 Installation of equipment
29
CLV-190 INSTRUCTION MANUAL
Ch.3
CAUTION
Do not place any equipment other than the video system center on the top of the
light source. Equipment damage can result.
Keep the ventilation grills of the light source clear. The ventilation grills are located
on the rear panels. Blockage can cause overheating and equipment damage.
Clean and vacuum dust the ventilation grills using a vacuum cleaner. Otherwise,
the light source may break down and gets damaged from overheating.
Place the light source on a stable, level surface using the foot holders (MAJ-1205).
Otherwise, the light source may topple down or drop, and operator or patient injury
may occur, or equipment damage can result.
When using the mobile workstation, confirm that excessive force is not applied to
the cord of the workstation. The cord may be disconnected, and malfunction can
result.
If a trolley other than the mobile workstation (WM-NP2, WM-DP2, WM-NP1,
WM-WP1, or WM-DP1) is used, confirm that the trolley can withstand the weight of
the equipment installed on it.
Do not install the light source in locations exposed to strong electromagnetic
radiation (e.g., in the vicinity of a microwave therapeutic equipment, MRI,
short-wave therapeutic equipment, radio equipment, or cellular/portable phones).
Otherwise, the light source may malfunction.
3.3 Installation of equipment

Table of Contents

Other manuals for Olympus EVIS EXERA III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Olympus EVIS EXERA III and is the answer not in the manual?

Olympus EVIS EXERA III Specifications

General IconGeneral
ModelEVIS EXERA III
CategoryMedical Equipment
TypeEndoscopy System
ManufacturerOlympus
System TypeVideo Endoscopy System
Video ProcessorCV-190
HD OutputYes
Power Requirements100-240 V AC, 50/60 Hz
Image QualityHDTV
Compatible EndoscopesEVIS EXERA III series endoscopes
Image EnhancementNBI (Narrow Band Imaging)
Light SourceXenon Lamp

Related product manuals