23
Maintenance
Maintenance Inspection and Safety Management
The HBP-1100 must be maintained to ensure functionality and to secure
the safety of patients and operators.
Daily checks and maintenance should be performed by the operator.In
addition, quali¿ ed personnel are necessary to maintain the performance
and the safety, and to conduct periodic inspections. We recommend that
the veri¿ cation test be performed at least once a year.
Monitor Maintenance
Cleaning and disinfecting should be performed in accordance with your
facility's infection control practice.
Wipe with a cloth that has been moistened with isopropyl alcohol diluted
to 50 v/v%, or ethyl alcohol (disinfection alcohol) diluted to 76.9 - 81.4
v/v% and wrung out.
Do not wipe the Power connector or allow it to become wet.
Use a moistened cotton bud to remove dust that has accumulated on the
vent ports.
The monitor requires no routine service other than cleaning, and visually
checking the cuffs, tubing, etc.
Caution
• Do not sterilize by autoclave or gas sterilization (EOG, formaldehyde
gas, high-concentration ozone, etc.).
• If using an antiseptic solution for cleaning, follow the instructions of the
manufacturer.
Accessory Care
Non-Invasive Blood Pressure Measurement (NIBP)
Cuff/Cuff tube
Wipe clean on the surface of the cuff with a cloth moistened with a
70 v/v% dilution of isopropyl alcohol, or a 76.9 to 81.4 v/v% dilution of
disinfection ethanol (ethyl alcohol).
Do not allow any liquids inside the cuff. If a liquid gets in the cuff, dry the
inside well.
Managing Supplies
Keep the following supplies on hand.
• GS CUFF - All sizes
• Battery