36
Manufacturer's Declaration
The HBP-1100 is intended for use in the electromagnetic environment
speci¿ ed below.
The customer or the user of the HBP-1100 should assure that it is used in
such an environment.
Electromagnetic Emissions (IEC60601-1-2)
Emission Test Compliance Electromagnetic Environment
RF emission
CISPR 11
Group 1
The HBP-1100 uses RF energy only for
internal functions. Therefore, this RF
emission is extremely weak and there is little
chance of it creating any kind of interference
whatsoever with nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The HBP-1100 is suitable for use in
all establishments, including domestic
establishments and those directly connected
to the public low voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage À uctuations/
À icker
IEC 61000-3-3
Complies