2
Introduction
Intended Use
Medical Purpose
This instrument is a digital monitor intended for use in measuring blood
pressure and pulse rate in adult and pediatric patient population with arm
circumference ranging from 12 cm to 50 cm (from 5 inches to 20 inches).
Intended User
This instrument should be used by a medical professional.
Patient Population
This instrument is intended for use on adults and children of age 3-years
and older.
Environment
The instrument is designed for use in physicians' of¿ ces, hospitals, clinics
and other medical facilities.
Measurement Parameter
Ŷ Non-Invasive Blood Pressure
Ŷ Pulse rate
Precautions for Use
Warnings and cautions described in the instruction manual should be
observed at all times.
Features of the Monitor
The blood pressure accuracy of the HBP-1100 is clinically proven. Easy
to use, the HBP-1100 is intended for use by medical professionals.
Ŷ Zero indicator function (page 17): Before each measurement, this
monitor indicates that "zero calibration" was successful.
Ŷ Auscultation mode
Ŷ 5 cuffs available - (XL: 42 to 50 cm, L: 32 to 42 cm, M: 22 to 32 cm, S:
17 to 22 cm, SS: 12 to 18 cm)
Ŷ This monitor and cuff can be cleaned with a soft cloth moistened with
alcohol.
Ŷ Compact, can be stored in a drawer
Ŷ Motion stop function: When body movement is detected, this monitor
stops deÀ ation for 5 seconds.
Ŷ Irregular pulse icon: Helps identify changes in heart rate, rhythm, and
pulse that may be caused by heart disease or other serious health
problems.