- 85 -
Stability marker
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
11 characters - right justification (digit markers contained)
Unit
3 characters - left justification
Example:
_ _ _ _1 8 . 3 4 3 [ 2 0 ] _ g _ _ CR LF - a printout generated, for <GLP printout template> settings as
given in the above table, from a weighing device upon pressing
key.
14.1.1. Response Format
On receipt of a command, the balance responds as follows.
first:
XX_A CR LF
command understood and in progress
XX_I CR LF
command understood but not accessible at this moment
XX _ ^ CR LF
command understood but max threshold is exceeded
XX _ v CR LF
command understood but min threshold is exceeded
XX _ E CR LF
time limit exceeded while waiting for stable measurement result
(time limit is balance characteristic parameter)
Where XX stands for command name.
next:
1 – 3
4
5
6
7
8 16
17
18 - 20
21
22
command
space
stability
marker
space
character
mass
space
unit
CR
LF
Command
3 characters
Stability marker
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
Character
[space] for positive values
[-] for negative values
Mass
9 characters - right justification
Unit
3 characters - left justification