repligen.com IF.UG.028 R1
Figure 3. Installed flow path pressure sensor ferrite bead
To install the supplied flow path pressure sensor ferrite bead:
1. Locate the ferrite bead as close as possible to the flow path pressure sensor while ensuring
there is enough cable to wrap around the ferrite bead.
2. Open the ferrite bead.
3. Seat the flow path pressure sensor cable in the bead.
4. Wrap the cable around the bead.
5. Snap the bead closed.