XCell
TM
ATF System with C410:V4B Controller User Guide
105
The XCell™ ATF should be in a Remote Control Mode.
Remote Set-up
The master sends input command and set up parameters to XCell™ ATF or set up the
parameters locally (on the HMI). The Start parameter should be “0”.
Command parameters are highlighted in light blue in the OPC inputs Table after this section.
For the XCell™ ATF controller to accept these parameters, the “CommandApply” bit must be
set. Similarly, setup parameters are highlighted in light red and the “ConfigurationSave” bit
must be set.
Remote Start
The master must send at least values in PFLOWSP, EFLOWSP, PRV1, PRV2, PV1 and PV2
(command parameters) to XCell™ ATF. The CommandApply bit must be set to “1” for these
values to be accepted by the XCell™ ATF controller. Initially, the Start and Stop bits should be
reset “0”. The master starts the filtration by setting the Start bit to “1”. The ATF starts running.
The Start bit can be reset to “0” after the ATF_Running bit is set and received from the XCell™
ATF controller.
Remote Stop
The master stops the filtration by setting the Stop bit to “1”. The ATF stops running. The Stop
bit can be reset to “0” after the ATF_Running bit is reset and received from the XCell™ ATF
controller.
Remote running
During a run it is possible to change all parameters remotely.
If the XCell™ ATF is changed from Remote Control Mode to Local Control Mode, it continues to
run, and it is possible to stop the run from the HMI. If the filtration has been stopped locally
and then changed back to Remote Control Mode, it will not restart. To restart the filtration the
master has to perform the items listed in “Remote Start” described earlier in this section.
System-stop
If the XCell™ ATF is stopped by the System-stop (one of 2 blue buttons on either the P-Box or E-
Box), and the System-stop is reset, the XCell™ ATF will not restart. The master has to perform
the items listed in “Remote Start” described earlier in this section to restart the filtration.
System-stop is indicated visually with the buttons illuminated, via DL_WARN_REG[5] = true.
The DL_WARN_REG[5] is not reset until a new start sequence is performed as described in the
“Remote Start” section.
Calibration Process
When not running, calibration activities can be performed.
Calibration parameters and commands are found on the second OPC Inputs (continued) page.
The master sends enable, capture and accept commands as well as high and low scaling
parameters to perform calibration activities, or calibration can be performed via the HMI.
Repligen recommends having a service technician perform calibration activities. Once loop
calibration is initiated, it must complete the calibration process. There is currently no means to
disable an enabled calibration without completing the process via remote commands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF Series and is the answer not in the manual?

REPLIGEN XCell ATF Series Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals