Chapter 6 MATH and Measurement RIGOL
MSO1000Z/DS1000Z User’s Guide 6-13
from the analog signal on the source. The setting range is from 0 Div to
0.96 Div.
Press Auto Scale to enable or disable the auto scale function. When auto
scale is enabled, the instrument will adjust the vertical scale of the
operation results to the optimal value according to the current
configuration.
Note: Smooth is grayed out and disabled. It is only available for differential
operation.
You can also use HORIZONTAL POSITION and HORIZONTAL
SCALE to adjust the horizontal position and scale of the operation results.
"XOR" Operation
Perform logic "XOR" operation on the waveform voltage values of the specified
sources point by point and display the results. In operation, when the voltage value
of the source channel is greater than the threshold of the corresponding channel, it is
regarded as logic "1"; otherwise logic "0".
The results of logic XOR operation of two binary bits are as follows.
Table 6-4 Logic "XOR" Operation
A
B
A^B
0
0
0
0 1 1
1 0 1
1
1
0
Press MATH Math Operator to select "A^B":
Press Operation to enable or disable the "XOR" operation function.
Press SourceA and SourceB to select the desired channels (CH1-CH4 or
D0-D15).
If source A selects any channel in CH1-CH4, press Thre.A and use set
the threshold of source A in logic operation. If source A selects any channel
in D0-D15, Thre.A will be hidden automatically.
If source B selects any channel in CH1-CH4, press Thre.B and use set
the threshold of source B in logic operation. If source B selects any channel
in D0-D15, Thre.B will be hidden automatically.
Press Offset and use to adjust the vertical position of the operation results.
Press Scale and use to adjust the vertical scale of the operation results.
Press Scale Reset to adjust the vertical scale of the operation results to the
optimal value according to the current configuration.
Press Options to set the start and end points of the operation results, enable or
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals