Chapter 4 To Set the Sample System RIGOL
MSO1000Z/DS1000Z User’s Guide 4-7
For digital channels:
When 8 channels are enabled, the memory depths available include Auto,
12kPoints, 120kPoints, 1.2MPoints, 12MPoints and 24MPoints (optional).
When 16 channels are enabled, the memory depths available include Auto,
6kPoints, 60kPoints, 600kPoints, 6MPoints and 12MPoints (optional).
Note: In "Auto" mode, the oscilloscope selects the memory depth automatically
according to the current sample rate.
Antialiasing
At slower sweep speed, the sample rate is reduced and a dedicated display algorithm
can be used to minimize the possibility of aliasing.
Press Acquire Anti-aliasing to enable or disable the antialiasing function. By
default, antialiasing is disabled and waveform aliasing is more possible.