RIGOL Chapter 8 Protocol Decoding
8-2 MSO1000Z/DS1000Z User’s Guide
Parallel Decoding
Parallel bus consists of clock line and data line. As shown in the figure below, CLK is
the clock line, while Bit0 and Bit1 are the 0 bit and 1st bit on the data line respectively.
The oscilloscope will sample the channel data on the rising edge, falling edge or the
rising/falling edge of the clock and judge each data point (logic "1" or logic "0")
according to the preset threshold level.
Figure 8-1 Parallel Decoding
Press MATH Decode1 Decoder to select "Parallel" and open the parallel
decoding function menu.
1. Press Decode to turn on or off the decoding function.
2. Clock Line Setting (CLK)
Press CLK to select any channel (CH1-CH4 or D0-D15) as the clock channel.
If "OFF" is selected, no clock channel is set.
Press Edge to set the oscilloscope to sample the channel data on the rising
edge (
), falling edge ( ) or rising/falling edge ( ) of the clock. If
no clock channel is selected, the instrument will sample when the channel
data jumps during the decoding.
3. Digital Bus
Press BUS to select the digital bus for parallel decoding. This setting will
automatically modify the settings of Width, Bit X and CH, as shown in the table
on the next page.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS1104Z Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z Plus and is the answer not in the manual?

Rigol DS1104Z Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals