RIGOL Chapter 18 Specifications
18-8 MSO1000Z/DS1000Z User’s Guide
General Specifications
Probe Compensation Output
Frequency
1 kHz
Power Voltage 100 V-240 V, 45 Hz-440 Hz
Temperature
Range
Operating: 0 ℃ to +50 ℃
Non-operating: -40 ℃ to +70 ℃
Cooling Method Fan cooling
℃
℃
: ≤95% Relative Humidity
℃
℃
: ≤75% Relative Humidity
+40 ℃ to +50 ℃: ≤45% Relative Humidity
Altitude Operating: under 3,000 meters
Non-operating: under 15,000 meters
Physical Characteristics
Width×Height×Depth=313.1 mm×160.8 mm×122.4 mm
The recommended calibration interval is one year.
Regulatory Information
Compatibility
Execution standard EN 61326-1:2006 EN 61326-2-1:2006
Safety UL 61010-1:2004; CAN/CSA-C22.2 NO. 61010-1-2004;
EN 61010-1:2001; IEC 61010-1:2001
Note
[1]
: Typical.
Note
[2]
: Maximum value. 50ns, single-channel mode, dots display, auto memory depth.
Note
[3]
: Supporting legs and handle folded, knob height included.
Note
[4]
: Standard configuration.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com