RIGOL Chapter 1 Quick Start
1-8 MSO1000Z/DS1000Z User’s Guide
Probe Compensation
When the probes are used for the first time, you should compensate the probes to
make them match the input channels of the oscilloscope. Non-compensated or
poorly compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1. Perform steps 1, 2, 3 and 4 of "Function Inspection".
2. Check the displayed waveforms and compare them with the following figures.
Figure 1-9 Probe Compensation
3. Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the displayed waveform is as the "Perfectly
compensated" in the figure above.
Over compensated Perfectly compensated Under compensated