EasyManuals Logo
Home>Rigol>Test Equipment>DS4000 Series

Rigol DS4000 Series User Manual

Rigol DS4000 Series
393 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
RIGOL Chapter 4 Sample System Setting
4-8 MSO4000/DS4000 User’s Guide
Peak Detect
In this mode, the oscilloscope acquires the maximum and minimum values of the
signal within the sample interval to get the envelope of the signal or the narrow pulse
that might be lost. In this mode, signal confusion can be prevented but the noise
displayed would be larger.
In this mode, the oscilloscope can display all the pulses with pulse widths at least as
wide as the sample period.
High Resolution
This mode uses an over-sample technique to average the neighboring points of the
sample waveform. This reduces the random noise on the input signal, generates a
much smoother waveform on the screen and improves the vertical resolution. This is
generally used when the sample rate of the digital converter is greater than the
storage rate of the acquisition memory.
Note: The Average” and High Resmodes use different averaging methods. The
former uses Multi-sample Average and the latter uses Single-sample Average.
www.calcert.com sales@calcert.com1.800.544.2843
0
5
10
15
20
25
30

Table of Contents

Other manuals for Rigol DS4000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS4000 Series and is the answer not in the manual?

Rigol DS4000 Series Specifications

General IconGeneral
BrandRigol
ModelDS4000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals