EasyManuals Logo
Home>Rigol>Test Equipment>DS7024

Rigol DS7024 User Manual

Rigol DS7024
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #119 background imageLoading...
Page #119 background image
Chapter 5 To Trigger the Oscilloscope RIGOL
MSO7000/DS7000 User Guide 5-27
Note: In Window trigger, pressing Trigger LEVEL can quickly switch the
current level adjustment type.
Rotate the Trigger LEVEL knob to adjust the corresponding trigger level.
During the adjustment of trigger level, two trigger level lines appear on the screen,
and they move up and down with the change of the trigger level. At the same time,
the real-time trigger level and slope information is displayed at the lower-left corner
of the screen (as shown in the figure below, H indicates the upper limit of the trigger
level, L indicates the lower limit of the trigger level, and â–³ indicates the trigger level
deviation). When you stopping modifying the trigger level, the trigger level line and
the trigger level information at the lower-left corner of the screen disappear in about
2 s. The current trigger level deviation is displayed at the upper-right corner of the
screen.
Trigger Mode:
In the trigger control area (Trigger) on the front panel, press Mode to quickly switch
the current trigger mode. For details, refer to descriptions in "Trigger Mode".
Trigger Parameter Setting:
Set the trigger parameters (trigger holdoff and noise rejection) under this trigger
type. For details, refer to descriptions in "Trigger Holdoff" and "Noise
Rejection".
Delay Trigger
In Delay trigger, you need to set Source A and Source B. The oscilloscope triggers
when the time difference (â–³T) between the specified edges of Source A (Edge A) and
Source B (Edge B) meets the preset time limit, as shown in the figure below.
Note: Edge A and Edge B must be two neighboring edges.

Table of Contents

Other manuals for Rigol DS7024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS7024 and is the answer not in the manual?

Rigol DS7024 Specifications

General IconGeneral
BrandRigol
ModelDS7024
CategoryTest Equipment
LanguageEnglish

Related product manuals