Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-151
:SYSTem:GPIB
Syntax
:SYSTem:GPIB <adr>
:SYSTem:GPIB?
Description
Sets or queries the GPIB address.
Parameter
Name
Type
Range
Default
<adr>
Integer
1 to 30
1
Return
Format
The query returns an integer ranging from 1 to 30.
Example
:SYSTem:GPIB 2 /*Sets the GPIB address to 2.*/
:SYSTem:GPIB? /*The query returns 2.*/
:SYSTem:KEY:PRESs
Syntax
:SYSTem:KEY:PRESs <key>
Description
Press down the specified key on the front panel of the oscilloscope.
Parameter
Name
Type
Range
Default
<key>
Discrete
{CH1|CH2|CH3|CH4|MATH|REF|LA|DECode|MOF
F|F1|F2|F3|F4|F5|F6|F7|NPRevious|NNEXt|NSTo
p|VOFFset1|VOFFset2|VOFFset3|VOFFset4|VSCa
le1|VSCale2|VSCale3|VSCale4|HSCale|HPOSition
|KFUNction|TLEVel|TMENu|TMODe|DEFault|CLE
ar|AUTO|RSTop|SINGle|QUICk|MEASure|ACQuir
e|STORage|CURSor|DISPlay|UTILity|FORCe|GEN
erator1|GENerator2|BACK|TOUCh|ZOOM|SEARc
h|WSCale|WPOSition}
——
Remarks
The value of the parameter <key> corresponds to the specified key on the front panel.
The definitions for the key are as follows:
 CH1|CH2|CH3|CH4|MATH|REF|LA|DECode|DEFault|CLEar|AUTO|SINGle|QUICk|ME
ASure|ACQuire|STORage|CURSor|DISPlay|UTILity|FORCe|BACK|TOUCh|ZOOM|SEA
Rch: press the specified key on the front panel.
 MOFF: hides or display the menu at the right side of the screen.
 F1|F2|F3|F4|F5|F6|F7: press the 7 menu softkeys at the right side of the screen.
 NPRevious|NNEXt|NSTop: press the navigation key.
 VOFFset1: press down the CH1 vertical offset knob, i.g. sets the vertical offset of
CH1 to 0.
 VOFFset2: press down the CH2 vertical offset knob, i.g. sets the vertical offset of
CH2 to 0.
 VOFFset3: press down the CH3 vertical offset knob, i.g. sets the vertical offset of
CH3 to 0.
 VOFFset4: press down the CH4 vertical offset knob, i.g. sets the vertical offset of
CH4 to 0.
 VSCale1: press down the vertical scale knob of CH1, i.g. switch the vertical scale of
CH1 to "Coarse" or "Fine".
 VSCale2: press down the vertical scale knob of CH2, i.g. switch the vertical scale of
CH2 to "Coarse" or "Fine".

Table of Contents

Other manuals for Rigol MSO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7000 Series and is the answer not in the manual?

Rigol MSO7000 Series Specifications

General IconGeneral
BrandRigol
ModelMSO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals