Page #233 background image
Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-219
:TRIGger:LIN (Option)
Command List:
:TRIGger:LIN:SOURce
:TRIGger:LIN:ID
:TRIGger:LIN:BAUD
:TRIGger:LIN:STANdard
:TRIGger:LIN:SAMPlepoint
:TRIGger:LIN:WHEN
:TRIGger:LIN:LEVel
:TRIGger:LIN:SOURce
Syntax
:TRIGger:LIN:SOURce <source>
:TRIGger:LIN:SOURce?
Description
Sets or queries the trigger source of LIN trigger.
Parameter
Name
Type
Range
Default
<source>
Discrete
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D11|D1
2|D13|D14|D15|CHANnel1|CHANnel2|CHANnel3|
CHANnel4}
CHANnel1
Return
Format
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14, D15,
CHAN1, CHAN2, CHAN3, or CHAN4.
Example
:TRIGger:LIN:SOURce CHANnel2 /*Sets the trigger source to CH2.*/
:TRIGger:LIN:SOURce? /*The query returns CHAN2.*/
:TRIGger:LIN:ID
Syntax
:TRIGger:LIN:ID <id>
:TRIGger:LIN:ID?
Description
Sets or queries the ID value of LIN trigger.
Parameter
Name
Type
Range
Default
<id>
Real
0 to 63
0
Return
Format
The query returns an integer ranging from 0 to 63.
Example
:TRIGger:LIN:ID 4 /*Sets the ID value of LIN trigger to 4.*/
:TRIGger:LIN:ID? /*The query returns 4.*/

Table of Contents

Other manuals for Rigol MSO7000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol MSO7000 Series and is the answer not in the manual?

Rigol MSO7000 Series Specifications

General IconGeneral
BrandRigol
ModelMSO7000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals