RIGOL Chapter 2 Command System
2-196 MSO7000/DS7000 Programming Guide
:TRIGger:NEDGe
Command List:
:TRIGger:NEDGe:SOURce
:TRIGger:NEDGe:SLOPe
:TRIGger:NEDGe:IDLE
:TRIGger:NEDGe:EDGE
:TRIGger:NEDGe:LEVel
:TRIGger:NEDGe:SOURce
:TRIGger:NEDGe:SOURce <source>
:TRIGger:NEDGe:SOURce?
Sets or queries the trigger source of the Nth Edge trigger.
{D0|D1|D2|D3|D4|D5|D6|D7|D8|D9|D10|D
11|D12|D13|D14|D15|CHANnel1|CHANnel2|
CHANnel3|CHANnel4}
The query returns D0, D1, D2, D3, D4, D5, D6, D7, D8, D9, D10, D11, D12, D13, D14,
D15, CHAN1, CHAN2, CHAN3, or CHAN4.
:TRIGger:NEDGe:SOURce CHANnel2 /*Sets the trigger source to CH2.*/
:TRIGger:NEDGe:SOURce? /*The query returns CHAN2.*/
:TRIGger:NEDGe:SLOPe
:TRIGger:NEDGe:SLOPe <slope>
:TRIGger:NEDGe:SLOPe?
Sets or queries the edge type of the Nth Edge trigger.
POSitive: indicates that the oscilloscope triggers on the rising edge of the input
signal when the voltage level meets the preset trigger level.
NEGative: indicates that the oscilloscope triggers on the falling edge of the input
signal when the voltage level meets the preset trigger level.
The query returns POS or NEG.
:TRIGger:NEDGe:SLOPe NEGative /*Sets the edge type to NEGative.*/
:TRIGger:NEDGe:SLOPe? /*The query returns NEG.*/