Roche/Hitachi cobas c 311 analyzer Communication Text
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 C-2
Communication Text
Table C-1 shows the communication text of the cobas c 311
analyzer.
Table C-1 List of Communication Text
cobas type
Text Direction
Real Time Batch
Inquiry for Order
Upload
cobas c 311->HOST
Yes No
Order for Test Request
Download
HOST-> cobas c 311
Yes Yes
Result Report
Upload
cobas c 311 ->HOST
Yes Yes
Inquiry of Result
Download
HOST-> cobas c 311
No Yes
(Yes: equipped, No: not equipped)
Note: Patient sample and quality control sample and
calibration result are sent as result data.
Note: There is no Inquiry for Order of a control sample and
Calibration.
Table C-2 shows Causes of Communication Text.
Table C-2 Causes of Communication Text
Text Real/Batch Cause
Inquiry for the
Requested Tests
(Upload)
Real
- After sample ID is read, inquire of HOST for test selection information of
patient sample to which test selection information is not registered.
- Wait for reply from HOST for test selection information for a certain length
of time after the inquiry. If not replied even after a certain length of time,
cancel the inquiry.
Real
- Specify the test selection information for a sample when the test selection
information is inquired.
Order for Test Request
(Download)
Batch
- HOST specifies the test selection information of a patient sample at a
given timing. Register test selection information before reading sample ID
to use this function.
Real
- Send result data of patient sample and quality control sample when test
data of the sample is collected.
Result Report
(Upload)
Batch
- Send result data of patient sample and quality control sample specified on
[Data Review] window.
Inquiry for result Batch
- Send result data of patient sample requested by Host. It is not possible to
select result from Host if 1st and rerun result exists, because the
selection is done at the Analyzer on Test Review screen. (Workplace /
Data Review / Test Review-Test result list).
The result is selected by the rules below from (1) to (3).
(1) There is a result.
(2) The result does not have alarm, which is listed on Review by
Exception screen.
(3) Result is selected if 1st and rerun result is existed in Data Review
screen or the only one result exists for a test.