Roche/Hitachi cobas c 311 analyzer Record structure
Photometric Absorbance Data Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-21
Photometric Absorbance Data Record
cobas type (Upload)
M|1|ABS|P1|1||10|50|0\1497\1499\1499|13140\12828\12760\12699
(1)(2)(3)(4) (5) (6) (7) (8)
\12646\12605\12571\12545\12526\12504\12495\12481\12474\12472\124
65\12463\12459\12454\12453\12449\12448\12442\12594\12986\12993
\12992\12987\12987\12980\12982\12975\12973\12974\12970\12972\129
70\12969\12969\12965\12970\12966\12970\12971\12974\12972\12977\1
2976\12982
[CR]
(9)
Comments
No
Field
No.
Field Type Max
cobas Type
(1) 1 Record Type ID ST 1 Use ‘M’.
(2) 2 Sequence Number NM 6 This record appears as the next layer of the Message Header Record. The field value is ‘1’
for one record. To specify several records, sequence numbers starting from 1 are assigned.
(3) 3 Record Type Sub ID ST 3 Use ‘ABS’.
(4) 4 Module ID ST 4 ID of the analytical unit (module) the test belongs to. The ID gives the first character of the
module that performed the analysis.
Module
Description
P1 cobas c 311 analyzer Module
ISE1 ISE Test
Non Calculate Test or Not measured test
(5) 5 Cell No NM 3 Number of the Reaction Cell.
Input value: 1-66
6 In Out Information
NM 1
Field does not contain data.
(6) 7 Reaction Time NM 2 Reaction time (Unit: minutes). Input values 3 to 10.
(7) 8 Point Num NM 2 Reaction point number. Input values 1 to 57.
(8) 9 Cell Blank Data NM 6 Repeat field.
Cell blank absorbance. The difference between the main wavelength and sub-wavelength
absorbance data [(main wavelength absorbance data) - (sub-wavelength absorbance data)]
can be repeated up to 4 points.1st data is ‘0’ fixed.
(9) 10 Delta ABS Data NM 6 Repeat field.
The difference between the main wavelength and the sub-wavelength absorbance data
[(main wavelength absorbance data) - (sub-wavelength absorbance data)] repeated up to the
value indicated by Point Num.
.