Roche/Hitachi cobas c 311 analyzer Record structure
Test Order Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-12
Test Order Record
cobas type (Upload)
O|1| 000663|6^50002^002^^S1^SC|^^^10^3|R||20050705093416
(1)(2) (3) (4) (5) (6) (7)
||||N||||1|||||||20050705095504|||F
[CR]
(8) (9) (10) (11)
cobas type (Download)
O|1| 000663|6^50002^002^^S1^SC|^^^10^3|R||20050705093416
(1)(2) (3) (4) (5) (6) (7)
||||A||||1||||||||||O
[CR]
(8) (9) (11)
Comments
No
Field
No.
Field Type Max
cobas Type
(1) 1 Record Type ID ST 1 ‘O’ fixed.
(2) 2 Sequence Number NM 6 Indicates sequence No. Normally it is ‘1’
Indicates the sequence number of the Test Order Record at the current layer. This record is
in the layer following the Patient Information Record and is reset to 1 for each occurrence of a
new Patient Information Record. It is numbered consecutively; 1, 2, etc., for each occurrence
of this record.
(3) 3 Specimen ID ST 13
or
22
Indicates sample No. (Sample ID.)
For control sample, set name or lot No. of control on communication setting screen.
Note:
In S.No. mode, becomes patient comment
Control: Control name or Lot-Number
(4) 4 Instrument Specimen
ID
CM 22 Indicates ordered sample. Setting is as follows:
< SequenceNo>^<Rack ID>^<PositionNo>^ ^
<SampleType>^<ContainerType>
<SequenceNo> Type: NM Max: 5
indicates cobas c 311 analyzer. internal sequence No.
Confirmation number when samples are numbered in sequence in order of input. In case
of QC material, ‘Control Number * 1000+ Sequence No. sample’ is used.
ex. Control No. = 1, Sequence No. = 2 is written as 1002
In ID mode, the number is reserved.
In S. No. mode, the range is 1 – 60000.
Control No. range is 1-100.
Control Seq. No. range is 1-150
<Rack ID> Type: ST Max: 5
indicates carrier No. (Disk/Rack.)
At the c311 analyzer, Rack ID contains two kinds of information:
Type of sample - 2 digits
20: Calibration
30: Control
40: STAT
50: Routine
+ Disk Position - 3 digits
Note
Because Host protocol and text field is same as cobas6000, Rack ID field is used in
cobas c 311 analyzer communication text
<PositionNo> Type: NM Max: 3
indicates position No. in carrier.
Values 001 ~ 110
<SampleType> Type: ST Max: 2
indicates sample type.
S1: Serum / Plasma
S2: Urine
S3: CSF
S4: Suprnt
S5: Others
QC: Control
<ContainerType> Type: ST Max: 7
indicates type of sample cup.
‘SC’: Standerd cup.
‘MC’:Micro cup.
no value Priority is given to the information inputted on the instrument