Roche/Hitachi cobas c 311 analyzer Record structure
Result Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-16
(10) 14 Instrument
Identification
ST 4
Indicates the ID of the analytical unit (module) that performed the analysis.
Module
Description
P1 cobas c 311 analyzer Module
ISE11 ISE Test
Non Calculate Test or Not measured test
Note: cobas Type transmits calculated tests.
Qualitative result format for cobas c 311 analyzer tests can
be activated on the Utility/Application/Range screen.
Table E-2 Qualitative Analysis Transmission Data
Analytical Result Range Transmission Data
Analytical Result =< Qualitative Analysis Concentration 1
-2
Qualitative Analysis Concentration 1 < Analytical Result =< Qualitative Analysis Concentration 2
-1
Qualitative Analysis Concentration 2 < Analytical Result =< Qualitative Analysis Concentration 3
0
Qualitative Analysis Concentration 3 < Analytical Result =< Qualitative Analysis Concentration 4
1
Qualitative Analysis Concentration 4 < Analytical Result =< Qualitative Analysis Concentration 5
2
Qualitative Analysis Concentration 5 < Analytical Result
3
Note: Communication of batch results is based on option of “with Review
by Exception Results” on [Send to Host] window shown below.
On: tests specified by Exception Alarm are communicated.
Off: tests specified by Exception Alarm are not communicated.