Roche/Hitachi cobas c 311 analyzer Record structure
Result Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-15
Result Record
cobas type (Upload)
R|1|^^^2/|8.60|nmol/L||N||F||BMSERV|20050912165312||P1[CR]
(1)(2) (3) (4) (5) (6) (7) (8) (9) (10)
Comments
No
Field
No.
Field Type Max
cobas Type
(1) 1 Record Type ID ST 1 ‘R’ fixed.
(2) 2 Sequence Number NM 6 Indicates sequence No.
Record in the current layer. Since this record is the layer that follows the Test Request
Record, it resets to 1 for each occurrence of a new Test Request Record. It is numbered
consecutively; 1, 2, etc., for each occurrence of this record.
(3) 3 Universal Test ID CM 24 Indicates order.
^^^<ApplicationCode>/<Dilution>/<pre-dilution>/…
<ApplicationCode> Type: NM Max: 5
indicates cobas c 311 analyzer Host Test No.
The analyzer identifies the test with 3-digit numbers. Specify these 3-digit numbers.
The range of application code is expanded to 5-digit.
Photometrics: 1-910, ISEs: Na=989, K=990, Cl=991
Serum Index: L=992, H=993, I=994, Calculated Tests: 961-968
<Dilution> Type: ST Max: 3
Indicates automatic dilution factor when ordering.
Inc, Dec,3,5,10,20,50
When not specified, pipetting and testing is done using the standard analysis parameters
(4) 4 Data or Measurement
Value
CM 13
Quantitative:
<measurement value>
Qualitative:
<qualitative value> ^< measurement value> or <qualitative value>
It depends on the setting in Host Communication Setting/Text Setting screen.
<measurement value> Type:NM Max:6
6-digit measurement value or 6 spaces if no result (or data has overflowed)
<qualitative value>
-2, -1, 0, 1, 2, 3 or 6 spaces if no result
In case that qualitative value is 0 and concentration is 8.60, example is as follows.
Example : 0^8.60
See Table E-2 Qualitative Analysis Transmission Data below for qualitative
(5) 5 Units ST 6 Indicates unit name of measurement results.
6 Reference Ranges CM
Field does not contain data.
(6) 7 Result Abnormal
Flags
ST 2 Indicates normal/abnormal of measurement results.
‘L’: less than normal range.
‘H’: more than normal range.
‘LL’: less than Technical Limit range.
‘HH’: more than Technical Limit range.
‘N’: Normal.
‘A’: Abnormal.
8 Nature of Abnormality
Testing
Field does not contain data.
(7) 9 Result Status ST 1 Indicates the number of the test conducted for the analytical data.
‘F’: initial result.
‘C’: rerun result.
10 Date of Change in
Instrument Normative
Values Units
Field does not contain data.
(8) 11 Operator Identification ST 6 Indicates operator ID who conducted measurement. HOST is not allowed to do setting.
(9) 12 Data/Time Test
Started
TS 14 Date and time corresponding the test was pipetted is designated by YYYYMMDDHHMMSS.
This data is sent to the host only when an option of “Send Time stamp of pipetting of Sample”
is selected.
13 Date/Time Test
Completed
TS 14
Field does not contain data.