Roche/Hitachi cobas c 311 analyzer Record structure
ISE Calibration Result Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-20
ISE Calibration Result Record
cobas type (Upload)
M|1|ICR|adm···|ISE11||||^^^^^^^|^^^^^^^|^^^^^^^|-32.4^-35.1^-28.1
(1)(2)(3) (4) (5)(6)(7)(8)(9) (10) (11) (12)
^-32.2^56.0^134.2^135.2^-0. 7|-34.9^-46.3^-25.8^-33.7^55.7^4.8^5
(13)
.04^-0.06|121.4^125.8^118.4^121.9^-42.0^102.0^99.5^-3.7
[CR]
(14)
Comments
No
Field
No.
Field Type Max
cobas Type
(1) 1 Record Type ID ST 1 Use ‘M’.
(2) 2 Sequence Number NM 6 This record appears as the next layer of the Message Header Record. The field value is ‘1’
for one record. To specify several records, a sequence number starting from 1 is assigned.
(3) 3 Record Type Sub ID ST 3 Use ‘ICR’.
(4) 4 Operator ID ST 6 ID of the operator that performed the calibration at the analyzer.
(5) 5 Module ID ST 5 ISExy
x : Module Number (1 only)
y : Calibration Type
‘1’ : Type A , ‘2’ : Type B
(6) 6 Na Calibration Alarm NM 3 Na test calibration alarm.
(7) 7 K Calibration Alarm NM 3 K test calibration alarm.
(8) 8 Cl Calibration Alarm NM 3 Cl test calibration alarm.
(9) 9 Na Data Alarm CM 31 Na test data alarm. (See Table E-3 Data Alarm List.)
<Internal standard solution electromotive force data alarm>^<Low solution electromotive
force data alarm>^<High solution electromotive force data alarm>^<Calibrator solution
electromotive force data alarm>^<Slope for display data alarm>^<Internal standard solution
concentration data alarm>^<Calibrator solution concentration data alarm>^<Numbers for
compensation data alarm>
Detail specification is shown below:
Element
Max Len Format
<Internal standard solution 3 NM
electromotive force data alarm>
<Low solution electromotive 3 NM
force data alarm>
<High solution electromotive 3 NM
force data alarm>
<Calibrator solution electromotive 3 NM
force data alarm>
<Slope for display data alarm> 3 NM
<Internal standard solution 3 NM
concentration data alarm>
<Calibrator solution concentration 3 NM
data alarm>
<Numbers for compensation data 3 NM
alarm>
(10) 10 K Data Alarm CM 31 K test data alarm. Refer to Na data alarm.
(See Table E-3 Data Alarm List.)
(11) 11 Cl Data Alarm CM 31 CI test data alarm. Refer to Na data alarm.
(See Table E-3 Data Alarm List.)
(12) 12 Na Data CM 55 Na test calibration analytical data.
(All Elements: Max Length = 6, Type =NM)
<Internal standard solution electromotive force data>^<Low solution electromotive force
data>^<High solution electromotive force data>^<Calibrator solution electromotive force
data>^<Slope for display data>^<Internal standard solution concentration data>^<Calibrator
solution concentration data>^<Numbers for compensations data>
(13) 13 K Data CM 55 K test calibration analytical data. Refer to Na data.
(14) 14 Cl Data CM 55 Cl test calibration analytical data. Refer to Na data.