Roche/Hitachi cobas c 311 analyzer Record structure
Photometric Calibration Result Record
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-19
(12) 12
Expired Flag
NM 1 Expired Flag
0:Calibration performed with “Not Expired Bottle”
1:Calibration performed with “Expired Bottle”
This data is sent to the host only when an option of “Chemistry Test Calibration Result
Detail” is selected.
(13) 13
Calibrator lot number
NM 8 1 to 99999999
This data is sent to the host only when an option of “Chemistry Test Calibration Result Detail”
is selected.
(14) 14
Pipetting Date and
Time
ST 14 Date and time corresponded the calibrator was scheduled measuring is designated by
YYYYMMDDHHMMSS.
This data is sent to the host only when an option of “Chemistry Test Calibration Result Detail”
is selected.
.