Roche/Hitachi cobas c 311 analyzer Record structure
Syntax
Roche Diagnostics May 2011
Host Interface Manual – Version 1.3 E-3
Communication sequences (message flow) of communication
message by application layer are shown below.
(1) Order for Test Request : real-time
cobas c 311 analyzer [HOST]
Inquiry for the Requested Tests
(TSREQ^REAL)
->
<-
Order for Test Request (TSDWN^REPLY)
(2) Order for Test Request : batch
[cobas c 311 analyzer] [HOST]
<-
Order for Test Request (TSDWN^BATCH)
(3) Result Report : real-time
[cobas c 311 analyzer] [HOST]
Result Report (RSUPL^REAL) ->
(4) Result Report : batch
[cobas c 311 analyzer] [HOST]
Result Report (RSUPL^BATCH) ->
(5) Inquiry of Result : real-time
[cobas c 311 analyzer] [HOST]
<- Inquiry of Result (RSREQ^ REAL)
Result Report (RSUPL^REAL) ->
(6) Photometric Calibration Result Report : real-time
[cobas c 311 analyzer] [HOST]
Photometric Calibration Result
Report (RCUPL^REAL)
->
(7) ISE Calibration Result Report : real-time
[cobas c 311 analyzer] [HOST]
ISE Calibration Result Report
(ICUPL^REAL)
->
(8) Photometric Raw Data Report : Batch
[cobas c 311 analyzer] [HOST]
Photometric Raw Data Report
(ABUPL^BATCH)
->