A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-88
3.6.2.5.3 GIBSON
To calculate LV volume by calculating the cube of a given region.
Measurement items: LVIDd, LVIDs;
Calculation items: EDV, ESV, SV, CO, EF, SI, CI, MVCF, FS
Use the formula:
EDV=
6
π
× (0.98×LVIDd+5.90) ×LVIDd
2
ESV=
6
π
× (0.98×LVIDs+5.90) ×LVIDs
2
Measurement method:
1. In M or B/M mode frozen state, press theCalckey to pop up the
Meas & Calc menu in M mode, move the trackball, use the cursor to
select “LV function”, press the SETkey to go to the subdirectory,
move the trackball, use the cursor to select “GIBSON”, press the
SETkey to confirm going to GIBSON measurement mode.
2. Measure LVIDd, see “3.6.3.1 Distance Measurement” for the specific
measurement.
3. Measure LVIDs, see “3.6.3.1 Distance Measurement” for the specific
measurement.
4. The measurement and calculation result is displayed in the result
window.
5. Repeat steps 1 to 4 to perform the next measurement.
3.6.3.6 Mitral Valve measurement (MV)
Function: In M or B/M mode frozen state, based on image from values for
points A to F, to navigate from points A~F, measure point by point, display
Amplitude of E wave (CE), Amplitude of DE wave (DE), Amplitude of A wave
(CA), Mitral value opening speed (DE), Mirral value closing speed (EF), Ratio
between amplitude A and amplitude E (A/E).