8.1.3.3 To Test IV Pole Height Adjustment
From the control screen menu, touch and hold the IV Pole
UP and DOWN arrow buttons to raise and lower the IV
pole, respectively (Figure 21). Make sure the height of
the IV pole increases and decreases correctly. If not, see
Section 13 Troubleshooting.
Figure 21 – To Test the Powered IV Pole
8.1.3.4 To Test Smoke Evacuation
NOTES:
Make sure a smoke evacuator ULPA filter is installed
in the rover before testing smoke evacuation. See
Section 5.3 For Use With. See the instructions for use
supplied with the smoke evacuator filter for installation
information.
See Section 8.1.4 To Adjust the Rover Settings to reset
the hours of use of a newly installed smoke evacuator
filter.
The rover monitors the use and life expectancy of the
smoke evacuator filter. DO NOT relocate the smoke
evacuator filter after initial installation into the rover.
See Table 14 Smoke Evacuator Modes of Operation for
mode descriptions. Make sure the ON or AUTO mode
is selected.
Table 14 – Smoke Evacuator Modes of Operation
MODE DESCRIPTION
OFF Stops smoke evacuator operation.
ON Evacuates continuously at the specified power
setting whether smoke is present or not.
AUTO Increases smoke evacuation to the specified
power setting when smoke is detected.
If smoke is not detected, the smoke
evacuator will operate at a 10% power setting
automatically.
1. Touch the EVACUATE SMOKE button. The EVACUATE
SMOKE dialog will appear (Figure 22).
2. While in the ON mode, touch the UP or DOWN buttons
to increase or decrease the smoke evacuator power
setting, as required.
3. Verify smoke evacuation power can be adjusted
correctly. If not, see Section 13 Troubleshooting.
4. After testing, touch the OFF button to turn off the
smoke evacuator; then touch the CLOSE button.
Set up system
Close
Power 60 %
Off On Auto
Evacuate
Smoke
Figure 22 – Evacuate Smoke Dialog
0000160552, Rev. C Effective Date: Nov 25, 2015 1:43:15 PM
Print Date: Nov 25, 2015 02:19:13 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker Neptune 3 and is the answer not in the manual?

Stryker Neptune 3 Specifications

General IconGeneral
BrandStryker
ModelNeptune 3
CategoryWater Pump
LanguageEnglish

Related product manuals