Appendix
C:
Performance Verification
instrument that is operating outside this temperature range. All tolerances
specified are for the instrument only and do not include test-equipment
error.
Test Equipment
Required
Table A-7 lists all the test equipment required to do the Performance Verifica-
tion in this appendix. Also listed is the minimum specifications for the test
equipment. All equipment used must meet or exceed its minimum specifica-
tions.
When equipment other than that recommended is used, control settings of
the test setup might need to be altered. If the exact item of equipment given
as an example in Table A-7 is not available, check the Minimum Specification
column to determine if any other available test equipment might suffice to
perform the check or adjustment.
Operating instructions for test equipment are not given in this procedure.
If
more operating information is required, refer to the appropriate test equip-
ment instruction manual.
Table
A-7:
Test Equipment Required
Item and Minimum Example of Suitable
Description Specification Purpose Test Equipment
Calibration Standard-amplitude signal levels: 5 mV to 50 V. Signal source for TEKTRONIX PG
506A
Generator Accuracy +-0.3%. gain and transient Calibration Genera-
High-amplitude signal levels: 1 V to 60 V. response. tor.'
Repetition rate: 1 kHz.
Fast-rise signal level: 1
V.
Repetition rate: 1 MHz.
Rise time: 1 ns or less. Flatness:
?2%.
Leveled Frequency: 250 kHz to above 100 MHz.
Sine Wave Output amplitude: variable from 10
mV to
5V
p-p.
Generator Output impedance: 50
Q.
Reference frequency: 50 kHz.
Amplitude accuracy: constant within 3% of refer-
ence frequency as output frequency changes.
Vertical, horizon- TEKTRONIX SG 503
tal, and triggering
Leveled Sine Wave
checks and ad- Generator.'
justments. Display
adjustments and
Z-Axis check.
Time-Mark
Marker outputs: 10 ns to 0.5 s.
Generator
Marker accuracy:
+
0.1
%.
Trigger output: 1 ms to 0.1 ms, time-coincident
with markers.
Horizontal checks TEKTRONIX TG 501
and adjustments. Time-Mark Genera-
Display adjust- tor.'
ment.
Low- Range: 1 kHz to 500 kHz. Low-frequency TEKTRONIX SG 502
4
~scillator.~
!!?
Frequency Output amplitude: 300 mV. trigger checks.
Generator Output impedance: 600 W.
4
Reference frequency: constant within 0.3
dB
of
Lr.
reference frequency as output frequency
-
changes.
5
pp
-
-
-
-
-
-
-
-
4
Pulse Repetition rate: 1 kHz. Output amplitude: 5 V. External clock and TEKTRONIX PG 501
tr.
Generator storage checks. Pulse Generator.'
.-.I
*
*r
A-26
Appendices
2
b