Appendix
C:
Performance Verification
i.
CHECK
-
Trace shift of 0.5 division or less.
j.
Press in the
SAVEICONT
button to select
CONT
k. Position the trace 0.5 division above the bottom horizontal graticule
line using the
CH
2
POSITION
control.
I.
Press in the
SAVEICONT
button to select
SAVE.
m. CHECK
-
Trace shift of 0.5 division or less.
n. Press in the
SAVEICONT
button to select
CONT
o. Set the Vertical
MODE
switch to
CH
1
p. Repeat steps g through m for
CH 1
trace.
Step
6:
Check Bandwidth
a. Set:
VOLTSIDIV
(both)
2
mV
AC-GND-DC
(both)
DC
A
SECIDIV
0.2 ms
BW LIMIT
Off
(button out)
STOREINON-STORE
NON-STORE
(button
out)
b. Connect the leveled sine wave generator output via a 50
fl
precision
cable and a 50
fl
termination to the
CH
1
OR
X
input connector.
c. Set the generator to produce a 50 kHz, six division display.
d. CHECK
-
Display amplitude is
4.2
divisions or greater as the gener-
ator output frequency is increased up to the value shown in Table
A-1 0 for the corresponding
VOLTSIDIV
switch setting.
Table A-1
0:
Settings for Bandwidth Checks
VOLTSIDIV
Generator
Switch Setting Output Frequency
2
mV
80 MHz
5
mV to 0.5 V 100 MHz
e. Repeat parts c and d for all indicated
CH 1 VOLTSIDIV
switch
settings, up to the output-voltage upper limit of the sine wave gener-
ator being used.
f.
Move the cable from the
CH 1
OR
X
input connector to the
CH
2
OR
Y
input connector.
g. Set the Vertical
MODE
switch to
CH
2.
h. Repeat parts c and d for all indicated
CH
2
VOLTSIDIV
switch
settings, up to the output-voltage upper limit of the sine wave gener-
ator being used.