Appendix
C:
Performance Verification
k. Set the second cursor on the trailing edge of the
1
oth
time marker at
the same voltage level as on the
2nd time marker.
I.
CHECK
-
The
AT
readout is between 0.798 ms and 0.802 ms.
m. Press in the
SAVEICONT
button to select
CONT
n. Set the
A SECIDIV
switch to 0.5 ps.
o. Select 0.5
pstime markers from the time-mark generator.
p. Align the
2"d time marker with the 2nd vertical graticule line using the
Horizontal
POSITION
control.
NOTE
Allow the points to accumulate for a few seconds before saving the
display.
q.
Repeat parts h through k.
NOTE
Pulses with fast rise and fall times have only a few sample points
and it may not be possible to place the cursors at exactly the same
voltage levels.
r.
CHECK
-
The
AT
readout is between 3.990 ps and 4.01 0 ps.
Step
3:
Check Variable Range and Sweep Separation
a. Set:
A
and
B SECIDIV
SECIDIV
Variable
STOREINON-STORE
0.2 ms
Fully counterclockwise
NON-STORE
(button
out)
b. Select 0.5 ms time markers from the time-mark generator.
c. CHECK
-
Time markers are one division or less apart.
d. Set:
CH
1
AC-GND-DC
GND
SECIDIV
Variable
CAL
detent
Horizontal
MODE
BOTH
B
TRIG
Fully clockwise (Runs
After Delay)
e. Use the
CH
1
POSITION
control to set the A Sweep at the center
horizontal graticule line.
f. CHECK
-
The
B
Sweep can be positioned more than 3.5 divisions
above and below the A Sweep when the
A16 SWP SEP
control is
rotated fully clockwise and counterclockwise respectively.
2232