EXAMPLE: The horizontal difference
is
0.6 division
with a sweep rate of 45° /division
as
shown
in
Figure 21.
Substituting the
given
values:
Phase
0 6 d' · · X 45°/d" · · 27°
Difference
= .
IVISJons
IVJslon
=
HIGH RESOLUTION PHASE DIFFERENCE
More
accurate phase measurements can
be
made
by
using the X10
MAG
mode to increase the sweep rate
without changing the
VAR
TIME/DIV control
(see
Figure 22).
EXAMPLE:
If
the sweep rate were increased 10 times
with the magnifier (X10
MAG),
the magnified sweep rate
would
be
45 + 10 = 4.5° /division. Figure 22 shows the same
signals used
in
Figure 21, but the X10
MAG
push button
is
depressed, resulting
in
a horizontal difference of 6
divisions. The phase difference
is:
Phase
Difference
horizontal
difference
(divisions)
magnified
X sweep rate
(degrees/div)
Substituting the
given
values:
Phase Difference
= 6 divisions X 4.5° /division = 27°