4.
Adjust
the
DELAY TIME POSITION control
to
move
the
reference
point
to
desired
spot
on
the
reference
signal trace (see Figure
33,
Point A).
5. Observe leading edge
of
the
8 Sweep reference signal
and use DELAY TIME POSITION control
to
move it
to
a convenient vertical graticule line.
6.
Adjust
the
b.
TIME control
to
move
the
time
measurement
point
to
the
desired
spot
on
the
Channel 2 trace (see Figure
33,
Point
8).
7. Observe
the
leading edge
of
the 8 Sweep comparison
signal
and
use
the
b.
TIME control
to
superimpose
the
reference and comparison signal leading edges.
8. Read
the
time difference on
the
DM44 Readout.