AFG1000 Series Specifications and Performance Verification
Frequency/Period test
This test verifies the frequency accuracy of the instrument. All output frequencies
are derived from a single generated frequency. Only one frequency point of
channel 1 is required to be checked.
1.
Connect the arbitrary function generator to the frequency counter as shown in
the following figure.
Figure 2: Frequency/Period tests
2.
Push the following buttons to recall the arbitrary function generator
default
setup:
Utility > System > Set to Default.
3.
Set up the arbitrary function generator as follows:
a.
Push the Sine button on the front panel.
b.
Press the Freq/Period bezel button to choose Freq. The chosen parameter
will be lighted with white background.
c.
Use the numeric keypad or the general purpose knob to set the
frequency to
1 MHz.
d.
Press the Ampl/High bezel button to choose Ampl. The chosen parameter
will be lighted with white background.
e.
Use the numeric keypad or the general purpose knob to set the
amplitude to 1.000 V
p-p
.
f.
Check that the CH1 On/Off front panel button LED is on. If not, then
the channel output is off. Push the CH1 On/Off button
to turn it on.
4.
Check that reading of the Frequency Counter is between 0.999999 MHz
and
1.000001 MHz.