Command Groups
Calibration a
nd Diagnostic Command Group
The Calibration and Diagnostic commands provide information about the current
state of o scilloscope calibration. They also initiate internal signal path calibration
(SPC)orexec
ute diagnostic tests. Commands that are specictofactory
calibration are not described in this manual. They are described in the Service
manual, available for download at www.tektronix.com.
Table 2-16: Calibration and Diagnostic C ommands
Command
Description
*CAL? Instructs the oscilloscope to perform
self-calibration and returns the oscilloscope
self calibration status
CALibrate:FACtory Provides the controls for starting and
stopping the factory calibration process
CALibrate:FACtory:STATus? Returns the factory calibration status value
saved in nonvolatile memory
CALibrate:INTERNal Starts a signal path compensation
CALibrate:INTERNal:STARt Starts the internal signal path calibration
CALibrate:INTERNal:STATus? Returns the current status of the internal
signal path calibration
CALibrate:RESults? Returns the status of all calibration
subsystems without performing an SPC
operation
CALibrate:RESults:FACtory? Returns the status of internal and factory
calibration
CALibrate:RESults:SPC? Returns the results of the last SPC operation
CALibrate:TEMPerature? Returns 0 as MSO/DPO2000/B series
does not support recording of oscilloscope
temperature
DIAg:LOOP:OPTion Sets the self-test loop option
DIAg:LOOP:OPTion:NTIMes Sets the self-test loop option to run N times
DIAg:LOOP:STOP Stops the self-test at the end of the current
loop
DIAg:RESUlt:FLAg? Returns the pass/fail status from the last
self-test sequence execution
DIAg:RESUlt:LOG? Returns the internal results log from the last
self-test sequence execution
DIAg:SELect:<function> Selects one of the available self-test areas
DIAg:STATE Sets the oscilloscope operating state
DIAg:SELect Runs self tests on the specied system
subsystem
2-16 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2000 Series and is the answer not in the manual?

Tektronix DPO2000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals