Analyze Wavefor
mData
Analyze Wavef
orm Data
After having properly set up the acquisition, triggering, and display of your desired waveform, you can then analyze the
results. Select from features such as cursors, automatic measurements, math, and FFT.
Taking Automatic Measurements
To take an automatic measurement:
1. Push Measure.
2. Push Add Measurement.
Add Mea-
surement
Remove
Measure-
ment
Indicators
Gating
Screen
High-Low
Method
Auto
Bring
Cursors on
Screen
Configure
Cursors
3. Turn multipurpose knob a to select the
specific measurement. If needed, then
turn multipurpose k nob b to select the
channel to measure on. Then push OK Add
Measurement.
4. To remove a measurement, push Remove
Measurement. Then push the side-bezel
menu f
or the measurement you want to
remove, or Remove All. Then push OK
Remove Measurem ent.
Quick Tips
To remove all measurements, select Remove All Measurements.
A symbol appears i nstead of the expected numerical measurement if a v ertical clipping condition exists. Part of the
waveform is above or below the display. To obtain a proper numerical measurement, turn the vertical scale and position
knobs to make all of the waveform appear in the display.
MSO2000B and DPO2000B Series Oscilloscopes User Manual 89