Installation
3. Connect the TPP0200/TPP0100 probe
tip and reference lead to the PROBE
COMP connecto
rs on the oscilloscope.
4. Push Default Setup.
5. Push Autoset. The screen should now
display a square wave, approximately
5Vat1kHz.
NOTE. For best performance, it is
recommended that you set the Vertical scale
to1V.
If the signal appears but is misshapen,
perform the procedures for compensating
the probe. (See page 12.)
If no signal appears, rerun the procedure.
If this does not remedy the situation,
have the oscilloscope serviced by
qualied service personnel.
Compensating a Passive Voltage Probe
When
ever you attach a passive voltage probe for the rst time to any input channel, compensate the probe to match i t to
the corresponding oscilloscope input channel.
To properly compensate your passive probe:
1. Follow the steps for the functional
check. (See page 11 .)
2. Check the shape of the displayed
waveform to determine if your
probe is properly compensated.
Properly compensated
Under compensated Over compensated
12 MSO2000B and DPO2000B Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix DPO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2024B and is the answer not in the manual?

Tektronix DPO2024B Specifications

General IconGeneral
BrandTektronix
ModelDPO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals