Acquisition
Using Autoset
Use Autoset to quickly and automatically set up the instrument (acquisition, horizontal, trigger, and vertical) based on the
characteristics of the input signal. Autoset makes adjustments to the signal such that the waveform displays two or three
cycles with the trigger level near the midlevel.
1. Connect the probe, and then select the
input channel. (See page 33, Setting up
analog signal input.)
2. Push the AUTOSET button to execute
an Autoset.
MSO/DPO70000DX, MSO/DPO7 0000C, and DPO7000C Series
MSO5000 and DPO5000 Series
3. Click Undo to undo the last Autoset.
Parameters that are not affected by
Autoset retain their settings.
Quick tips
Aut
oset automatically sets up the analog channels.
On i
nstruments with iCapture, autoset automatically sets up the iCapture channels.
36 MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO /DP O5000 Series U ser Manual