Commands Listed in Alphabetical Order
AFG:PERIod
Sets (or queries) the period of the AFG waveform, in seconds. (Period =
1.0/Frequency).
The period value is constrained by the frequency max/min/resolution (see
AFG:FREQuency). For example, frequency resolution steps from 0.1Hz to 1Hz,
result in th
e following valid settings for period: 10, 5, 3.33…, 2.5, 2, 1.66…,
1.428571…, 1.25, 1.11…, 1.0.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:PERIod <NR3>
AFG:PERIod?
Arguments
Floating point number that represents the AFG period value, in seconds.
Returns
The query response is returned in high precision NR3 format (up to 12 digits with
more than 4 trailing 0 digits after the decimal point is omitted).
Examples
AFG:PERIOD 1 sets the AFG period value to 1 second.
AFG:PHASe
Sets (or queries) the AFG phase. The AFG phase setting controls the phase
difference between the trigger signal output and the AFG waveform output.
Phase is expressed in degrees and ranges from -180.0 to 180.0 in increments of
0.1 degrees.
Conditions
MDO3000 series with option MDO3AFG installed and MDO4000C series with
option MDO4AFG installed only.
Group
AFG
Syntax
AFG:PHASe <NR3>
AFG:PHASe?
2-140 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual