Commands Listed in Alphabetical Order
SAVEWFM, MASK:
TESt:SAVEIMAGE, MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,orMASK:TESt:SRQ:FAILure.
Aseriesofexa
mples showing how to use mask commands for typical tasks
is included in an appendix. (See page E-1, Mask/Limit Command Sequence
Examples.)
Conditions
This command requires a DPO4LMT application module for 4000 series models,
and an MDO3LMT application module for MDO3000 series models.
Group
Mask
Syntax
MASK:STOPOnviolation {ON|OFF|<NR1>}
MASK:STOPOnviolation?
Related Commands
MASK:TESt:SAVEWFM,
MASK:TESt:SAVEIMAGE,
MASK:TESt:AUXout:FAILure,
MASK:TESt:HARDCopy,
MASK:TESt:SRQ:FAILure
Arguments
ON stops waveform acquisition
on the first occurrence of a mask violation.
OFF turns this feature off.
<NR1>=0 turns this feature off ; any other value turns it on.
Examples
MASK:STOPONVIOLATION ON sets the instrument to stop waveform acquisition
on the first occurrence of a waveform violation.
MASK:STOPONVIOLATION? might return :MASK :STOPONVIOLATION ON,
indicating that the waveform acquisitions will be stopped upon the first occurrence
of a waveform violation.
MASK:TEMPLate:CREATEmask (No Query Form)
This command, used for limit mask testing, causes a template mask to be created
based on the settings of the MASK:TEMPLate:SOUrce, MASK:TEMPLate:
TOLerance:HORizontal,andMASK:TEMPLate:TOLerance:VERTical
commands. If you do not specify horizontal or vertical tolerances, the default
tolerance of one minor division will be used.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-381