EasyManuals Logo
Home>Tektronix>Test Equipment>MDO4000B Series

Tektronix MDO4000B Series User Manual

Tektronix MDO4000B Series
1158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1104 background imageLoading...
Page #1104 background image
Appendix E: Mask/Limit Command Sequence Examples
Command Comment
:MASK:TEST:SAVEIMAGE O N Once the test starts, this command will save a .PNG, .BMP or .TIF le (depending upon the
setting of SAVe:IMAGe:FILEFormat) of the waveform from Channel 1 upon mask violation to a
le in the curr
ent working directory.
:MASK:TEST:
STATE 1
This starts the test.
Next, force
a mask violation. Generate a 300KHz sine wave, 0.22 volt Peak to peak, 0.0 offset.
:ACQUIRE:S
TATE?
This query should return 0, indicating that the acquisitions have stopped, because a violation
has taken pl
ace.
:FILESYST
EM:DIR?
This query
should show that a new waveform and a new screen image le have been saved to
the current working directory.
E-4 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals