Commands Listed in Alphabetical Order
if the position
for Channel 3 is set to 2.0 and the offset is set to 3.0, then input
signals equal to 3.0 units are displayed 2.0 divisions above the center of the screen
(at 1 V/div).
Group
Vertical
Syntax
CH<x>:POSition <NR3>
CH<x>:POSition?
Related Commands
CH<x>:OFFSet, REF<x>:VERTical:POSition, MATH[1]:VERTical:POSition
Arguments
<NR3> is a floating point number that specifies the position value for channel <x>,
in divisions, from the center graticule. The range is 8 to -8 divisions.
Examples
CH2:POSition 1.3 positions the Channel 2 input signal 1.3 divisions above the
center graticule.
CH1:POSition? might return :CH1:POSITION -1.3000 indicating that the
current position of Channel 1 is 1.3 divisions below the center graticule.
CH<x>:PRObe? (Query Only)
Returns all information concerning the probe attached to channel <x>, where x
is the channel number.
Group
Vertical
Syntax
CH<x>:PRObe?
CH<x>:PRObe:AUTOZero (No Query Form)
This command executes the attached probe’s Auto Zero function, for probes that
support this feature. See your probe documentation for more details.
Group
Vertical
Syntax
CH<x>:PRObe:AUTOZero EXECute
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-225