Commands Listed in Alphabetical Order
MATH specifies t
hat the Math waveform data will be transferred from the
oscilloscope to the controller.
REF1–REF4 spe
cifies which Reference waveform data will be transferred from
the oscilloscope to the controller.
D0–D15 spec
ifies which digital channel waveform data will be transferred from
the oscilloscope to the controller. (MSO/MDO4000/B models only as well as
MDO3000 and MDO400C models with option MDO3MSO or MDO4MSO
installed.)
DIGital specifies that the Digital Collection waveform data will be transferred
from the oscilloscope to the controller. (MSO/MDO4000/B models only as well
as MDO3000 and MDO400C models with option MDO3MSO or MDO4MSO
installed.)
RF_AMPlitude|RF_FREQuency|RF_PHASe (MDO4000/B/C models only)
and
RF_NORMal|RF_AVErage| RF_MAXHold|RF_MINHold specify that the
RF data will be transferred from the oscilloscope to the controller (MDO models
only).
Examples
DATA:SOURCE? might return :DATA:SOURCE REF3 indicating that Reference
location 3 is the source for the waveform data that is transferred using a
CURVe?
query.
DATA:SOURCE CH1 specifies that the CH1 waveform will be transferred in the
next CURVe? query.
DATa:STARt
This command specifies the starting data point for incoming or outgoing
waveform transfer using the
CURVe command or query. (Use DATa:STOP to
specify the ending data point.) You can set the
DATa:STARt and DATa:STOP
values automatically to the starting and stopping points of the waveform cursors,
if on, using
DATa SNap. Note that setting DATa:STARt and DATa:STOP
automatically sets WFMOutpre:NR_Pt.
(See page 2-95, Waveform Transfer Command Group.)
Group
Waveform Transfer
Syntax
DATa:STARt <NR1>
DATa:STARt?
Related Commands
CURVe,
2-278 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual