Commands Listed in Alphabetical Order
Group
Math
Syntax
MATH[1]:VERTical:POSition <NR3>
MATH[1]:VERTical:POSition?
Related Commands
CH<x>:POSition, REF<x>:VERTical:POSition
Arguments
<NR3> is a floating point number that spec
ifies the desired position in divisions
from the center graticule.
Examples
MATH1:VERTICAL:POSITION 1.3E+00 positions the Math 1 input signal
1.3 divisions higher than a position of 0.
MATH1:VERTICAL:POSITION? might return :MATH1:VERTICAL:POSITION
-1.3000E+00
indicating that the current position of Math 1 is 1.3 divisions
below the center graticule.
MATH[1]:VERTical:SCAle
This command specifies the vertical scale of the currently selected math type. This
setting controls the display only, graphically scaling these waveforms and having
no affect on the acquisition hardware. For a signal with constant amplitude,
increasing the scale causes the waveform to be displayed smaller. Decreasing the
scale causes the waveform to be displayed larger.
Group
Math
Syntax
MATH[1]:VERTical:SCAle <NR3>
MATH[1]:VERTical:SCAle?
Related Commands
CH<x>:SCAle,
REF<x>:VERTical:SCAle
Arguments
<NR3> is a floating point number that specifies the scale-per-division in the current
math vertical units. The range is from 1.0E-12 through 500.0E+12.
Examples
MATH1:VERTICAL:SCALE 100E-03 sets the Math scale to 100 mV per division.
MATH:VERTICAL:SCALE? might return :MATH:VERTICAL:SCALE
1.0000E+00
indicating that the current scale setting of Math is 1 V per division.
2-418 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual