Commands Listed in Alphabetical Order
Examples
RF:SQUELCH:ST
ATE ON
turns the squelch control on.
RF:SQUELCH:STATE? might return 0, indicating that squelch has been turned off.
RF:SQUELCH:THReshold
This command specifies the squelch threshold level, in volts, for the RF Frequency
vs. Time and RF Phase vs. Time traces.
Conditions
MDO4000/B/C models only.
Group
RF
Syntax
RF:SQUELCH:THReshold <NR3>
RF:SQUELCH:THReshold?
Related Commands
RF:SQUELCH:STATE
Arguments
<NR3> is a floating point number.
Examples
RF:SQUELCH:THReshold 200.0000E-6 sets the squelch threshold level to
200 μV.
RF:SQUELCH:THReshold? might return 22.0000E-6, indicating the squelch
threshold level is 22 μV.
RF:STARt
This command specifies to exclude frequencies below a certain level from use.
NOTE. For the parameters RF:STARt, RF:STOP, RF:FREQuency and RF:SPAN,
when any one of the parameters is changed, the following relationship is enforced:
START = FREQ — SPAN/2
STOP = FREQ + SPAN/2
FREQ = (START + STOP)/2
SPAN = STOP — START
Conditions
MDO3000, MDO4000/B, or MDO4000C models with option SA3 or SA6 only.
2-588 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual