Commands Listed in Alphabetical Order
Examples
TRIGGER:A:VID
EO?
might return :TRIGGER:A:VIDEO:STANDAR D
BILEVELCUSTOM;SYNC ALLLINES;LINE 4;HOLDOFF:FIELD
0.0E+0;:TRIGGER:A:VIDEO:POLARITY POSITIVE;CUSTO M:FORMAT
PROGRESSIVE;LINEPERIOD 63.5600E-6;SYNCINTERVAL 4.7200E-6
TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe}
This command sets the video trigger format (either interlaced or progressive)
to use for triggering on video signals. To use this command, you must also
set the video standard to
BILevelcustom or TRILevelcustom (using
TRIGger:A:VIDeo:STANdard).
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe}
{INTERLAced|PROGressive}
TRIGger:A:VIDeo:CUSTom{:FORMat|:TYPe}?
Arguments
INTERLAced argument sets the format to interlaced video lines.
PROGressive argument
sets the format to progressive video lines.
Examples
TRIGGER:A:VIDEO:CUSTOM:FORMAT PROGRESSIVE sets the custom format for
the A video trigger t
o progressive lines.
TRIGGER:A:VIDEO:CUSTOM:FORMAT? might return :TRIGGER:A:VIDEO:
CUSTOM:FORMAT INTERLACED
indicating that interlaced is selected as the
custom format for the A video trigger.
TRIGger:A:VIDeo:CUSTo m:LINEPeriod
This command sets the line period for the standard under test. To use this
command, you must also set the video standard to
BILevelcustom or
TRILevelcustom (using TRIGger:A:VIDeo:STANdard).
Conditions
This command requires a DPO4VID application module to use any standard
besides NTSc, PAL, or SECAM, except for MDO3000 models, which do not
require an application module.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-913