Use the Normal mode when you want to see only valid triggered waveforms.
When you use this mode, the oscilloscope does not display a waveform until after
the first trigger.
To perform a Single Sequence acquisition, push the Single button.
Source Options.
Source option Details
Ch1, Ch2 Triggers on a channel whether or not the waveform is displayed
Ext Does not display the trigger signal; the Ext option uses the signal
connected to the Ext Trig front-panel BNC and allows a trigger level
range of +1.6 V to -1.6 V
Ext/5 Same as Ext option, but attenuates the signal by a factor of five, and
allows a trigger level range of +8 V to -8 V; this extends the trigger level
range
AC Line
1
Uses a signal derived from the power line as the trigger source; trigger
coupling is set to DC and the trigger level to 0 volts.
You can use the AC Line when you need to analyze signals related to the
frequency of the power line, such as lighting equipment and power supply
devices; the oscilloscope automatically generates the trigger, sets the
Trigger Coupling to DC, and sets the Trigger Level to zero volts.
NOTE. To view an Ext, Ext/5, or AC Line trigger signal, push and hold down the
Trigger Menu button for more than 1.5 seconds to enable trigger view.
Coupling. Coupling allows you to filter the trigger signal used to trigger an
acquisition.
Option Details
DC Passes all components of the signal
Noise Reject Adds hysteresis to the trigger circuitry; this reduces sensitivity which
reduces the chance of falsely triggering on noise
HF Reject Attenuates the high-frequency components above 80 kHz
LF Reject Blocks the DC component and attenuates the low-frequency components
below 300 kHz
AC Blocks DC components and attenuates signals below 10 Hz
NOTE. Trigger coupling affects only the signal passed to the trigger system. It
does not affect the bandwidth or coupling of the signal displayed on the screen.
1
Available only when you select the Edge trigger type.
Reference
TBS1000B and TBS1000B-EDU Series Oscilloscopes User Manual 103

Table of Contents

Other manuals for Tektronix TBS1052B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TBS1052B and is the answer not in the manual?

Tektronix TBS1052B Specifications

General IconGeneral
BrandTektronix
ModelTBS1052B
CategoryTest Equipment
LanguageEnglish

Related product manuals