Performance Tests
4–46
TDS 500B, 600B and TDS 700A Service Manual
1. Install the test hookup and preset the instrument controls:
a. Initialize the instrument:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
H Press the side-menu button OK Confirm Factory Init.
b. Modify the default setup:
H Press VERTICAL MENU.
H Press the main-menu button Coupling; then press the side-menu
button to select 50 coupling.
H Set the horizontal SCALE to 10 ns on the TDS 600B and 12.5 ns on
the TDS 500B/700A.
c. Hook up the test-signal source: Connect the output of a
medium-frequency leveled sine wave generator (Item 25) to CH 1. Do
this through a 50 W precision coaxial cable, followed by a 10X
attenuator. See Figure 4–14.
50 Coaxial Cable
Medium-Frequency
Sine Wave Generator
Digitizing Oscilloscope
Output
10X Attenuator
Figure 4–14: Initial Test Hookup
2. Confirm the trigger system is within time-accuracy limits for pulse-glitch or
pulse-width triggering (Horizontal Scale 1 s):
a. Display the test signal: Set the output of the sine wave generator for a
100 MHz, five-division sine wave on screen. Press SET LEVEL TO
50%.

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals