Acquiring and Displaying Waveforms
3–10
TDS 500B, TDS 600B, & TDS 700A User Manual
When you turn off a waveform, the oscilloscope automatically selects the next
highest priority waveform. Figure 3–4 shows the order of priority.
1. CH1
2. CH2
3. CH3 or AX1
4. CH4 or AX2
1. MATH1
2. MATH2
3. MATH3
4. REF1
5. REF2
6. REF3
7. REF4
Figure 3–4: Waveform Selection Priority
Note Figure 3–4 shows two orders of priority due to the following rules: If you
are turning off more than one waveform and you start by turning off a channel
waveform, all channels will be turned off before going to the MORE waveforms.
If you start by turning off the MORE waveforms, all the MORE waveforms will
be turned off before going to the channel waveforms.
If you turn off a channel that is a trigger source, it continues to be the trigger
source even though the waveform is not displayed.
To read about selecting reference waveforms, see Saving and Recalling
Waveforms on page 3–118.
To read about selecting (and creating) math waveforms, see Waveform Math on
page 3–148.
Scaling and Positioning Waveforms
The TDS Oscilloscope allows you to scale (change the vertical or horizontal
size) and position (move up, down, left, or right) waveforms on screen for best
display. (Figure 3–5 shows the results of both vertical and horizontal scaling and
positioning.) This section first tells you how to quickly check and set vertical
and horizontal scales, positions, and other parameters, such as vertical bandwidth
and horizontal record length.
To quickly see the position of the waveform in the display, check the Channel
Reference and the Record View icons. (See Figure 3–5.)
The Channel Reference icon, at the left side of the display, points to ground on
the waveform record when offset is set to 0 V. The oscilloscope contracts or
expands the selected waveform around this point when you change the vertical
scale.
Waveform Priority
To Find More Information
To Check Position

Table of Contents

Other manuals for Tektronix TDS 700A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 700A and is the answer not in the manual?

Tektronix TDS 700A Specifications

General IconGeneral
BrandTektronix
ModelTDS 700A
CategoryTest Equipment
LanguageEnglish

Related product manuals