Performance Tests
TDS 500D, TDS 600C, TDS 700D & TDS 714L Performance Verification and Specifications
121
TDS 600C Test Record
Photocopy this and the next three pages and use them to record the performance
test results for your TDS 600C. The TDS 694C Test Record begins on
page 129.
TDS 600C Test Record
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 600C Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy
CH1 Offset +1 mV
+101 mV
+1.01 V
2.1 mV
75.6 mV
756 mV
__________
__________
__________
__________
__________
__________
+ 2.1 mV
+ 75.6 mV
+ 756 mV
CH2 Offset +1 mV
+101 mV
+1.01 V
2.1 mV
75.6 mV
756 mV
__________
__________
__________
__________
__________
__________
+ 2.1 mV
+ 75.6 mV
+ 756 mV
CH3 or AX1 +1 mV
Offset +101 mV
+1.01 V
2.1 mV
75.6 mV
756 mV
__________
__________
__________
__________
__________
__________
+ 2.1 mV
+ 75.6 mV
+ 756 mV
CH4 or AX2 +1 mV
Offset +101 mV
+1.01 V
2.1 mV
75.6 mV
756 mV
__________
__________
__________
__________
__________
__________
+ 2.1 mV
+ 75.6 mV
+ 756 mV
DC Voltage Measurement Accuracy (Averaged)
CH1 5 mV Vert scale setting,
5 Div position setting, +1 V offset
+ 1.0329 V __________ __________ + 1.0471 V
CH1 5 mV Vert scale setting,
+5 Div position setting, 1 V offset
1.0471 V __________ __________ 1.0329 V
CH1 200 mV Vert scale setting,
5 Div position setting, +10 V offset
+ 11.4165 V __________ __________ + 11.7835 V
CH1 200 mV Vert scale setting,
+5 Div position setting, 10 V offset
11.7835 V __________ __________ 11.4165 V
CH1 1.01 V Vert scale setting,
5 Div position setting, +10 V offset
+ 17.102 V __________ __________ + 18.899 V
CH1 1.01 V Vert scale setting,
+5 Div position setting, 10 V offset
18.899 V __________ __________ 17.102 V
CH2 5 mV Vert scale setting,
5 Div position setting, +1 V offset
+ 1.0329 V __________ __________ + 1.0471 V
CH2 5 mV Vert scale setting,
+5 Div position setting, 1 V offset
1.0471 V __________ __________ 1.0329 V
CH2 200 mV Vert scale setting,
5 Div position setting, +10 V offset
+ 11.4165 V __________ __________ + 11.7835 V
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Tektronix TDS 784D

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS 784D and is the answer not in the manual?

Tektronix TDS 784D Specifications

General IconGeneral
BrandTektronix
ModelTDS 784D
CategoryTest Equipment
LanguageEnglish

Related product manuals